Например, Бобцов

NOISE CANCELLATION IN UNSHIELDED MAGNETOCARDIOGRAPHY BASED ON LEAST-MEAN-SQUARED ALGORITHM AND GENETIC ALGORITHM

Аннотация:

This paper discusses adaptive noise cancellation in magnetocardiographic systems within unshielded environment using two algorithms, namely, the Least-Mean-Squared (LMS) algorithm and the Genetic Algorithm (GA). Simu- lation results show that the GA algorithm outperforms the LMS algorithm in extracting a weak heart signal from a much-stronger magnetic noise, with a signal-to-noise ratio (SNR) of -35.8 dB. The GA algorithm displays an improvement in SNR of 37.4 dB and completely suppresses the noise sources at 60Hz and at low frequencies; while the LMS algorithm exhibits an improvement in SNR of 33 dB and noisier spectrum at low frequencies. The GA algorithm is shown to be able to recover a heart signal with the QRS and T features being easily extracted. On the other hand, the LMS algorithm can also recover the input signal, however, with a lower SNR improvement and noisy QRS complex and T wave.

Ключевые слова:

Статьи в номере